Evaluation and routine testing in medical imaging departments-Acceptance and constancy tests. Imaging performance of computed tomography X-ray equipment
医学影像科的评估和常规检测
This part of IEC 61223 applies to CT SCANNERS that conform to IEC 60601-2-44:2009, IEC 60601-2-44:2009/AMD1:2012 and IEC 60601-2-44:2009/AMD2:2016.
IEC 60601-2-44 and this document
• defines the essential parameters which describe the performance of CT SCANNERS with regard to image quality, RADIATION OUTPUT and PATIENT positioning; the list of parameters to be tested can be found in 4.3,• defines the methods of testing the essential parameters, and• evaluates compliance with the tolerances of the parameters SPECIFIED by the ACCOMPANYING DOCUMENTS.
The methods defined in IEC 60601-2-44 and this document rely on non-invasive measurements, using appropriate test equipment, performed during or after installation. Signed statements covering steps in the installation procedure can be used as part of the ACCEPTANCE TEST report.Cross References:IEC/TR 60788:2004EN 60601-1:2006EN 60601-2-44:2009/A2:2016IEC 60601-2-44:2009/AMD1:2012IEC 60601-1:2005IEC TR 60788:2004EN 60601-1:2006/A12:2014EN 60601-2-44:2009/A11:2011IEC 60601-1:2005/AMD1:2012IEC 60601-2-44:2009EN 60601-1:2006/A1:2013EN 60601-2-44:2009/A1:2012EN 60601-2-44:2009IEC 60601-2-44:2009/AMD2:2016IEC 61674:2012All current amendments available at time of purchase are included with the purchase of this document.