Standard Test Method for Determination of Impurities in Plutonium: Acid Dissolution, Ion Exchange Matrix Separation, and Inductively Coupled Plasma-Atomic Emission Spectroscopic (ICP/AES) Analysis
钚杂质测定的标准试验方法:酸溶解、离子交换基质分离和电感耦合等离子体原子发射光谱(ICP/AES)分析
1.1
This test method covers the determination of 25 elements in plutonium (Pu) materials. The Pu is dissolved in acid, the Pu matrix is separated from the target impurities by an ion exchange separation, and the concentrations of the impurities are determined by inductively coupled plasma-atomic emission spectroscopy (ICP-AES).
1.2
This test method is specific for the determination of impurities in 8 M HNO
3
solutions. Impurities in other plutonium materials, including plutonium oxide samples, may be determined if they are appropriately dissolved (see Practice
C1168
) and converted to 8 M HNO
3
solutions.
1.3
The values stated in SI units are to be regarded as standard. The values given in parentheses are mathematical conversions that are provided for information only and are not considered standard. Additionally, the non-SI units of molarity and centimeters of mercury are to be regarded as standard.
1.4
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section
9
on Hazards.
1.5
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
====== Significance And Use ======
5.1
This test method can be used on plutonium matrices in nitrate solutions.
5.2
This test method has been validated for all elements listed in Test Methods
C757
except sulfur (S) and tantalum (Ta).
5.3
This test method has been validated for all of the cation elements measured in
Table 1
. Phosphorus (P) requires a vacuum or an inert gas purged optical path instrument.