Measurement microphones - Part 7: Values for the difference between free-field and pressure sensitivity levels of laboratory standard microphones
测量传声器第7部分:实验室标准传声器的自由场和压力灵敏度之间的差值
This part of IEC 61094
- gives a polynomial function derived from a least square fit to data from several laboratories, for the differences between free field and pressure sensitivity levels of laboratory standard microphones as specified in IEC 61094-1,
- enables determination of the free field sensitivity level of a laboratory standard microphone for zero degrees incidence in air by adding values of these differences to the pressure sensitivity level,
- gives tabulated values for the polynomial function for a range of frequency and temperature,
- is applicable when a suitable free field calibration is not available.