Standard Practice for Determining Relative Image Quality Response of Industrial Radiographic Imaging Systems below 4 MeV
4以下工业射线照相成像系统相对图像质量响应测定的标准实施规程 MeV
1.1
This standard covers a practice whereby industrial radiographic imaging systems or techniques may be comparatively assessed using the concept of relative image quality response (RIQR). Changes within a radiographic technique such as film/detector types, distances, or filtering/collimation can be comparatively assessed using this standard. The RIQR method presented within this practice is based upon the use of equivalent penetrameter sensitivity (EPS) described within Practice
E1025
and subsection
5.4
of this practice.
Fig. 1
illustrates a relative image quality indicator (RIQI) that has four different plaque thicknesses (0.38 mm, 0.25 mm, 0.20 mm, and 0.13 mm (0.015 in., 0.010 in., 0.008 in., and 0.005 in.)) sequentially positioned (from top to bottom) on an absorber plate of a specified material and thickness. The four plaques contain a total of 14 different arrays of penetrameter-type hole sizes designed to render varied conditions of threshold visibility when exposed to the appropriate radiation. Each “EPS” array consists of 30 identical holes; thus, providing the user with a quantity of threshold sensitivity levels suitable for relative image qualitative response comparisons. There are two standard materials (steel and plastic) specified herein for the RIQI and absorber. For special applications the user may design a non-standard RIQI-absorber configuration; however the RIQI configuration shall be controlled by a drawing similar to
Fig. 1
. Use of a non-standard RIQI-absorber configuration shall be described in the user’s written technique and approved by the CEO.
1.2
This practice is not intended to qualify the performance of a specific radiographic technique nor for assurance that a radiographic technique will detect specific discontinuities in a specimen undergoing radiographic examination.
1.3
This practice is not intended to be used to classify or derive performance classification categories for radiographic imaging systems. For example, performance classifications of radiographic film systems may be found within Test Method
E1815
, and manufacturer characterization of computed radiography (CR) systems may be found in Practice
E2446
. However, the RIQI and absorber described in this practice are used by Practice
E2446
for manufacturer characterization of computed radiography (CR) systems and by Practice
E2445
to evaluate performance and to monitor long term stability of CR systems.
1.4
These tests are for applications below 4 MeV. When a gamma source or other high energy source is used, these tests may still be used to characterize the system, but may need a modification of the absorber thickness to adjust the available RIQR range as agreed between the user and cognizant engineering organization (CEO). For high-energy X-ray applications (4 MV to 25 MV), Test Method
E1735
provides a similar RIQR standard practice.
1.5
The values stated in SI are to be regarded as the standard.
1.6
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
====== Significance And Use ======
4.1
This standard provides a practice for RIQR evaluations of film and non-film imaging systems when exposed through an absorber material. Three alternate data evaluation methods are provided in Section
9
. Determining RIQR requires the comparison of at least two radiographs or radiographic processes whereby the relative degree of image quality difference may be determined using the EPS plaque arrangement of
Fig. 1
as a relative image quality indicator (RIQI). In conjunction with the RIQI, a specified radiographic technique or method must be established and carefully controlled for each radiographic process. This practice is designed to allow the determination of subtle changes in EPS that may arise to radiographic imaging system performance levels resultant from process improvements/changes or change of equipment attributes. This practice does not address relative unsharpness of a radiographic imaging system as provided in Practice
E2002
. The common element with any relative comparison is the use of the same RIQI arrangement for both processes under evaluation.
4.2
In addition to the standard evaluation method described in Section
9
, there may be other techniques/methods in which the basic RIQR arrangement of
Fig. 1
might be utilized to perform specialized assessments of relative image quality performance. For example, other radiographic variables can be altered to facilitate evaluations provided these differences are known and documented for both processes. Where multiple radiographic process variables are evaluated, it is incumbent upon the user of this practice to control those normal process attributes to the degree suitable for the application. Specialized RIQR techniques may also be useful with micro focus X-ray, isotope sources of radiation or with the use of non-film radiographic imaging systems. RIQR may also be useful in evaluating imaging systems with alternate materials (RIQI and base plate) such as plastic, copper-nickel, or aluminum. When using any of these specialized applications, the specific method or techniques used shall be as specified and approved by the cognizant engineering organization.