1.1
This test method covers the measurement of alpha particle activity of water. It is applicable to nuclides that emit alpha particles with energies above 3.9 MeV and at activity levels above 0.02 Bq/mL (540 pCi/L) of radioactive homogeneous water. This test method is not applicable to samples containing alpha-emitting radionuclides that are volatile under conditions of the analysis.
1.2
This test method can be used for either absolute or relative determinations. In tracer work, the results may be expressed by comparison with a standard that is defined to be 100 %. For radioassay, data may be expressed in terms of alpha disintegration rates after calibration with a suitable standard. General information on radioactivity and measurement of radiation has been published in Refs
(
1-
3
)
2
and summarized in Practices
D3648
.
1.3
The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
====== Significance And Use ======
5.1
This test method was developed for the purpose of measuring gross alpha radioactivity in water. It is used for the analysis of both process and environmental water to determine gross alpha activity which is often a result of natural radioactivity present in minerals.