Standard Test Method for Determination of Low Mass Density/Fractions of Uranium in Oils, Organic Liquids, and Pellets by X-ray Fluorescence
用X射线荧光法测定油、有机液体和颗粒中低质量密度/分数铀的标准试验方法
1.1
This test method covers the steps necessary for the preparation and analysis by X-ray fluorescence (XRF) of oils and organic solutions containing uranium. Two different preparation techniques are described.
1.2
The procedure is valid for organic solutions with uranium mass densities of 20 μg/mL to 2000 μg/mL as presented to the spectrometer for the solution technique and pellets with uranium mass fractions of 200 μg/g to 50 000 μg/g for the pellet technique.
1.3
This test method requires the use of an appropriate internal standard. Care must be taken to ascertain that samples analyzed by this test method do not contain the internal standard or that this contamination, whenever present, has been corrected for mathematically. Such corrections are not addressed in this procedure. Care must be taken that the internal standard and sample medium are compatible; that is, samples must be miscible with tri-
n
-butyl phosphate (TBP) and must not remove the internal standard from solution. Alternatively, a scatter line may be used as the internal standard.
2
1.4
The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.5
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
Specific precautionary statements are given in Section
9
and
Note 2
.
1.6
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
====== Significance And Use ======
5.1
This test method is applicable to organic solutions containing uranium at 20 μg/mL to 2000 μg/mL presented to the spectrometer for the solution techniques or pellets containing uranium at 200 μg/g to 50 000 μg/g using the fused pellet technique.
5.2
Either wavelength-dispersive or energy-dispersive XRF systems may be used, provided that the software accompanying the system is able to accommodate the use of internal standards.