Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); System Reference Document for inductive Loop - Ultra Low Power Active Medical Implants (ULP- AMI) - systems operating in the frequency bands 9 kHz to 315 kHz
电磁兼容性和无线电频谱管理(ERM);短程设备(SRD);感应回路系统参考文件.超低功率有源医用植入物(ULP-AMI).在9 kHz至315 kHz频带内运行的系统
Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); System Reference Document for inductive Loop - Ultra Low Power Active Medical Implants (ULP- AMI) - systems operating in the frequency bands 9 kHz to 315 kHzThe present document applies to the radio sections of inductive loop Ultra Low Power Active Medical Implants in the
field of Short Range Devices (SRDs) transmitters and receiver systems operating on various frequencies within the
range of 9 kHz to 315 kHz.The present document gives guidance on the definition of a transmitter mask for the operating frequencies, analyses and
the spectrum requirements for transmission of power and data from/to stationary programmers or mobile
neurostimulation systems (pain control) to/from active medical implants.The present document considers the ERC Report 44, [1] and contains the technical characteristics for radio equipment
as given in CEPT/ERC Recommendation 70-03 [2]. Therefore, the frequency range 9 kHz to 135 kHz is already
sufficiently regulated for inductive systems in CEPT/ERC Recommendation 70-03 [2]. The necessary operation for the
frequency range 135 kHz to 315 kHz is not addressed in CEPT/ERC Recommendation 70-03 [2] and is addressed in the
present document.The spectrum parameters for inductive loop ULP-AMI devices that are the subject of this document are intended to be
included as a normative annex in the EN 300 330-2 [3] or in a new European Harmonized Standard. They are submitted
for consideration to CEPT/ERC WGSE for compliance studies and in WGFM/SRD MG for inclusion in the CEPT/ERC
Recommendation 70-03 [2], annex 12.