IEC 60942:2017 RLV contains the International Standard and its Redline version. The Redline version is available in English only. The Redline version provides you with a quick and easy way to compare all the changes between this standard and its previous edition. The Redline version is not an official IEC Standard, only the current version of the standard is to be considered the official document.
IEC 60942:2017 specifies the performance requirements for three classes of sound calibrator: class LS (Laboratory Standard), class 1 and class 2. Acceptance limits are smallest for class LS and greatest for class 2 instruments. Class LS sound calibrators are normally used only in the laboratory; class 1 and class 2 are considered as sound calibrators for field use. A class 1 sound calibrator is primarily intended for use with a class 1 sound level meter and a class 2 sound calibrator primarily with a class 2 sound level meter, as specified in IEC 61672-1. The acceptance limits for class LS sound calibrators are based on the use of a laboratory standard microphone, as specified in IEC 61094-1, for demonstrations of conformance to the requirements of this document. The acceptance limits for class 1 and class 2 sound calibrators are based on the use of a working standard microphone, as specified in IEC 61094-4, for demonstrations of conformance to the requirements of this document. This fourth edition cancels and replaces the third edition published in 2003, of which it constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) deletion of the class designations, class LS/C, class 1/C and class 2/C;
b) addition of two further class designations, class LS/M and class 1/M, specifically for pistonphones;
c) addition of an amended criterion for assessing conformance to a specification: conformance is now demonstrated when (a) measured deviations from design goals do not exceed the applicable acceptance limits and (b) the uncertainty of measurement does not exceed the corresponding maximum-permitted uncertainty;
d) modification to the short-term level fluctuation test of the sound pressure level stability;
e) change to some environmental test conditions to avoid icing;
f) addition of an alternative test for immunity to radio-frequency fields using transverse electromagnetic (TEM) waveguides.