Electromagnetic compatibility and Radio spectrum Matters (ERM); Short Range Devices (SRD); Medical Body Area Network Systems (MBANSs) operating in the 2 483,5 MHz to 2 500 MHz range; Part 1: Technical characteristics and test methods
电磁兼容性和无线电频谱管理(ERM);短程设备(SRD);在2483.5MHz至2500 MHz范围内运行的医疗机构区域网络系统(MBANSs);第1部分:技术特性和试验方法
发布日期:
2014-11-01
ETSI EN 303 203-1涵盖了医疗机构区域网络系统(MBANS)的最低特性,
包括频谱监控和接入要求,认为这是充分利用
在2 483,5 MHz至2 500 MHz频率范围内的可用频谱,并避免
该乐队的MBA和其他用户。它不一定包括一个项目可能需要的所有特征
它也不一定代表可实现的最佳性能。可以属于MBANS的设备类型包括体内和体外医疗传感器、患者监测设备
以及《医疗器械指令》(指令93/42/EEC[i.7])所涵盖的医疗致动器。本文件适用于以下被视为在室内操作的MBANS应用:在医疗设施中操作的MBANS在患者家中操作的MBANS本文件包含以下MBANS无线电设备的基本技术特征,这些特征也在CEPT/ERC/REC 70的附件2中讨论-
03[i.2]:医疗设施MBA,最大e.i.r.p.为1MW,最大工作周期不超过10%
发射带宽为3兆赫。最大e.i.r.p.为10 mW,最大工作循环不超过2%的患者家庭MBA
发射带宽为3兆赫。
ETSI EN 303 203-1 covers the minimum characteristics of Medical Body Area Network System (MBANS),
including the spectrum monitoring and access requirements, considered necessary in order to make the best use of the
available spectrum within the 2 483,5 MHz to2 500 MHz frequency range and to avoid harmful interference between
MBANS and other users of this band. It does not necessarily include all the characteristics which may be required by a
user, nor does it necessarily represent the optimum performance achievable.The types of devices that can belong to MBANSs are on-body and off-body medical sensors, patient monitoring devices
and medical actuators covered by the Medical Device Directive (Directive 93/42/EEC [i.7]).The present document applies to the following MBANS applications which are considered to operate indoor:MBANS operating in the healthcare facilityMBANS operating in the patient's homeThe present document contains the following basic technical characteristics of MBANS radio equipment which are also addressed in annex 2 of CEPT/ERC/REC 70-03 [i.2]:Healthcare facility MBANS with 1 mW maximum e.i.r.p. and not more than 10 % duty cycle over a maximum
emission bandwidth of 3 MHz.Patient's home MBANS with 10 mW maximum e.i.r.p. and not more than 2 % duty cycle over a maximum
emission bandwidth of 3 MHz.