Standard Test Method for Isotopic Abundance Analysis of Uranium Hexafluoride and Uranyl Nitrate Solutions by Multi-Collector, Inductively Coupled Plasma-Mass Spectrometry
通过多集电极 电感耦合等离子体质谱法对六氟化铀和硝酸铀溶液进行同位素丰度分析的标准测试方法
1.1
This test method covers the isotopic abundance analysis of
234
U,
235
U,
236
U, and
238
U in samples of hydrolysed uranium hexafluoride (UF
6
) by inductively coupled plasma source, multicollector, mass spectrometry (ICP-MC-MS). The method applies to material with
235
U abundance in the range of 0.2 to 6 % mass. This test method is also described in ASTM STP 1344.
1.2
The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
====== Significance And Use ======
5.1
The test method is capable of measuring uranium isotopic abundances of
234
U,
235
U,
236
U, and
238
U as required by Specifications
C787
and
C996
.