1.1 This practice covers procedures for testing and locating the sources of gas leaking at the rate of 2.2 10
14
mol/s (5 10
10
Std cm
3
/s). The test may be conducted on any device or component across which a pressure differential of halogen tracer gas may be created, and on which the effluent side of the area to be leak tested is accessible for probing with the halogen leak detector.
1.2 Five methods are described:
1.2.1 Method A - Direct probing with no significant halogen contamination in the atmosphere.
1.2.2 Method B - Direct probing with significant halogen contamination in the atmosphere.
1.2.3 Method C - Shroud test.
1.2.4 Method D - Air-curtain shroud test.
1.2.5 Method E Accumulation test.
1.3 The values stated in inch-pound units are to be regarded as the standard. The metric equivalents of inch-pound units may be approximate.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
====== Significance And Use ======
Halogen leak testing can be used to indicate the presence, location and magnitude of leaks in a closed vessel. This test method is normally used for production examination. Its use with halogenated refrigerant gases has been declining because of concerns about the effect of these gases on the ozone layer.