1.1
This test method provides procedures for validating viscosity measurements by rotational viscometers of Newtonian fluids. Performance parameters determined include viscosity repeatability (precision), detection limit, quantitation limit, linearity, and bias.
1.2
Validation of apparatus performance and analytical methods is requested or required for quality initiatives or where results may be used for legal purposes.
1.3
The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
====== Significance And Use ======
5.1
This test method may be used to validate the performance of a specific rotational viscometer apparatus.
5.2
This test method may be used to validate the performance of a specific method based upon the measurement of viscosity using rotational viscometer apparatus.
5.3
This test method may be used to determine the repeatability of a specific apparatus, operator, or laboratory.
5.4
This test method may be used for specification or regulatory compliance purposes.