Methods for the calibration of vibration and shock transducers — Part 17: Primary calibration by centrifuge
振动和冲击传感器的校准方法 - 第17部分:离心机初步校准
发布日期:
2016-05-30
ISO 16063包括一系列涉及振动和冲击传感器校准方法的文件。
ISO 16063-17:2016规定了使用离心机校准进行加速计初级校准的仪器和程序的详细规范。
ISO 16063-17:2016适用于零频率响应的直线加速度计,主要是应变计或压阻式加速度计,以及主要标准和工作传感器。
它适用于0 Hz下10 m/s2至20 000 m/s2(可能更高加速度)的校准范围。
适用的不确定度限值为读数的±1%。
ISO 16063 comprises a series of documents dealing with methods for the calibration of vibration and shock transducers.
ISO 16063-17:2016 lays down detailed specifications for the instrumentation and procedure to be used for primary calibration of accelerometers using centrifuge calibration.
ISO 16063-17:2016 is applicable to rectilinear accelerometers with zero-frequency response, mainly of the strain gauge or piezoresistive type, and to primary standard and working transducers.
It is applicable for a calibration range from 10 m/s2 to 20 000 m/s2 (higher accelerations possible) at 0 Hz.
The limits of uncertainty applicable are ±1 % of reading.