Standard Practice for Calibration and Functionality Checks Used in Forensic Psychophysiological Detection of Deception (Polygraph) Examinations
校准和功能检查的标准实践用于法医心理生理检测欺骗(测谎)考试
1.1 This practice provides guidelines for calibration and instrumentation (both analog and computerized systems) actively used in the psychophysiological detection of deception (
verification of truth
). As a minimum, such instrumentation shall simultaneously record an individual's respiratory, exosomatic electrodermal, and cardiovascular activity.
1.2 Analog polygraphs shall be calibrated by the psychophysiological detection of deception (PDD) examiner, manufacturer, or factory-authorized individual.
1.3 Computerized instrumentation shall be calibrated by the manufacturer or factory-authorized individual.
1.4 The PDD examiner or factory-authorized individual shall perform functionality checks to ensure instrumentation is operating properly.
1.5 This practice does not prohibit additional components which may be offered as supplemental measurements of physiological change. Additional recording components (such as movement sensors) may be used in addition to, but not replace, the required minimum components and these additional components shall meet the manufacturer's specifications for calibration.
====== Significance And Use ======
This practice sets forth the minimum requirements for calibration and functionality checks when conducting PDD examinations and related activities. For additional information see Practice
E1954
and Guide
E2000
.