1.1
This document describes the appearance of a radiographic image where fundamental components of image quality are changed, that is, variables such as whether an X-ray or gamma ray source was used, the characteristics of the radiographic detector, and specimen thickness (the geometrical configuration of the radiographic set-up and focal spot size both affect image quality but are not considered in the reference images).
1.2
The X-ray film systems used in obtaining the original illustrative data were as follows: Very Fine Grain (comparable to class I of Test Method
E1815
) and Fine Grain (comparable to class II of Test Method
E1815
).
1.3
These digital reference images consist of four composite illustrations and show how such factors as radiation energy, specimen thickness, and detector properties affect the radiographic image. The digital reference images are an adjunct to this document and must be purchased separately from ASTM if needed.
1.4
The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.
1.5
Only licensed copies of the software and images shall be utilized for production inspection. A copy of the ASTM/User license agreement shall be kept on file for audit purposes. (See
Note 1
.)
Note 1:
The set of digital reference images consists of four digital data files, and software to load the desired format and specific instructions on the loading process. The four reference images illustrate the effects of radiographic parameters in material from one inch to six inches thick. Available from ASTM International Headquarters, Order No: RRE3421.
1.6
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
====== Significance And Use ======
4.1
These reference images are intended as examples of how radiographic parameters affect the radiographic image. The use of these images in determining acceptance of production components is not intended or recommended.
4.2
A key consideration with any radiographic system is its contrast resolution and spatial resolution capability (that is, sensitivity). The degree of obtainable sensitivity with a given system is dependent upon several radiographic parameters such as source energy level, detector type, and material thickness radiographed. These reference images permit the user to estimate the degree of sensitivity change that may be obtained when these parameters are varied from a specific technique. This standard may also be used in conjunction with Test Method
E1815
or with Practice
E746
or Practice
E1735
to provide a basis for developing data for evaluation of a user's specific system. This data may assist a user in determining appropriate parameters for obtaining desired degrees of radiographic system sensitivity.